Why does the Pocket Colorimeter II™ display E2?
Hach Chlorine Pocket Colorimeter Ii
![Pocket Pocket](https://au.hach.com/cbf7d442c8a9ac8ad7a4074d4b14e634a3/cms/documents/zip/FAQ/faq-1.png)
Hach Pocket Colorimeter Ii Chlorine
The Hach Pocket Colorimeter II is calibrated for one to two parameters, with low or high range. Each instrument comes in a ready-to-use kit that includes a manual, pre-measured unit dose reagents, sample cells, and a sturdy custom carrying case. THE VERSATILE AND EASY-TO-USE POCKET COLORIMETER™ II RUGGED BACKLIT WATERPROOF. https://counterkeen487.weebly.com/user-manual-for-wondershare-video-converter-ultimate.html. View and Download Hach POCKET COLORIMETER II instruction manual online. https://counterkeen487.weebly.com/crestron-dmc-4k-c-hdcp2-user-manual.html. ANALYSIS SYSTEMS, Sulfate (SO4), Silica (SiO2). POCKET COLORIMETER II Measuring Instruments pdf manual download.
E2 Error on the Pocket Colorimeter II™
Hach Colorimeter Instrument
Remove power from the Pocket Colorimeter instrument by removing one of its batteries to verify that it is or is not the source of the interference. Move the Pocket Colorimeter instrument away from the device receiving the interference. Reposition the receiving antenna for the device receiving the interference. Try combinations of the.
Hach Pocket Colorimeter Ii Manual
![Hach pocket colorimeter ii expanded user manual software Hach pocket colorimeter ii expanded user manual software](/uploads/1/2/6/4/126482183/613718296.jpg)
Hach Pocket Colorimeter Ii User Manual
An E-2 error message on the Pocket Colorimeter II™ indicates the LED (light source) is out of regulation. Try replacing the batteries. Make sure that the LED in the cell holder comes on when the the read or zero buttons are pushed. If the LED doesn't come on, this indicates that the unit should be sent in for repair or replaced. This can also be related to a bad interference filter. Excess moisture can increase the failure of these filters.